SPECIAL CHROMATOGRAPHY WORKSTATION FOR SULFIDE ANALYSIS
For the detection of sulfide, the concentration is not proportional to the "peak height" but to the "square root of the peak height" (I. e., the so-called "exponential method" quantification). When selecting "single point correction method" (calibration factor needs to be calculated by standard sample first) or "multi-point correction method" (working curve needs to be calculated by standard sample first) for quantification, "square root of peak area" or "square root of peak height" can be selected for calculation. where the "square root of peak area" is actually the sum of the square roots of the slice heights that make up the peak.
Weight
0
Quantity
-
+
Inventory
0
隐藏域元素占位
- Product Description
-
- for the detection of sulfides, the concentration is not related toInstead, the "peak height" is proportional to the "square root of the peak height" (I. e., the so-called "exponential method" of quantification). When selecting "single point correction method" (calibration factor needs to be calculated by standard sample first) or "multi-point correction method" (working curve needs to be calculated by standard sample first) for quantification, "square root of peak area" or "square root of peak height" can be selected for calculation. where the "square root of peak area" is actually the sum of the square roots of the slice heights that make up the peak.
- When the working curve needs to be calculated, just.The "quantitative basis" in the "quantitative method table" is changed from the conventional "peak height" to the "square root of peak height" and then the working curve can be calculated, or the working curve with logarithmic horizontal and vertical axes can be calculated specifically.
- Usually used"Multi-point correction method (with multiple working curves)" to calculate the concentration of each component in turn to take a working curve, how many working curves can calculate how many components. If there is only one component (e. g. hydrogen sulfide) in the standard sample, and multiple similar sulfur components in the sample to be tested need to be quantified using its calculated working curve, you can select "multi-point correction (with a working curve)" in the "quantitative method table" for calculation.