Beijing Bei Fen Tianpu Instrument Technology Co., Ltd


Beijing Beifen Tianpu Instrument Technology Co., Ltd.

+
  • 二次热解析.jpg

Number:

1200124399726514176

Automatic secondary thermal analyzer/thermal desorption instrument


Classification:

Thermal desorption instrument Thermal desorption instrument
Secondary thermal analyzer

Hotline:

400-005-7070

Weight

0

Quantity
-
+

Inventory

0

隐藏域元素占位

  • Product Description
  • Parameter Description
  • Application areas
  • File Download
  • Automatic secondary thermal analyzer/Thermal desorption instrument       

    Model:ATD-12R/24R  

     

     

    Instrument features and main functions:

    The secondary thermal analyzer is fully automatic and one-button start-up, which automatically completes the desorption sample analysis process of all adsorption tubes without personnel on duty.

    Synchronous startGC/GCMChromatographic data processing workstation, or external event program can be used to start the device.,Improve efficiency and reduce sample waste

    Lower cold trap temperature and higher heating rate to ensure narrow chromatographic peak shape

    Automatic leak detection and fault alarm function

    Stable Servo Motor Drive Reliable Hardware and Software Control System

    Secondary thermal analyzer touch screen control, interface information rich, complete, simple operation

    Method parameter setting, real-time display of working status, running time

    Adsorption tube, injection valve, transfer tube, focusing tube (refrigeration, heating), five channels can be individually controlled temperature

    10This method is available for editing, storage and recall at any time, and the sample analysis is automatically completed by pressing the run key.

    High temperature six-way valve injection, greatly reducing the sample residue, to ensure the accuracy of the injection

    The connection point between the six-way valve and the transmission pipeline is in the heating insulation box, and there is no transmission cold point, which ensures the integrity of the sample.

    Automatic monitoring of the service life of the focusing tube of the secondary thermal analyzer.

    This machine has its own standard sample analog sampling function, which can make the working curve more conveniently through the thermal analyzer.

     

     

    The new 《GB 50325-2020Indoor Environmental Pollution Control Standards for Civil Building Engineering2020Year8Month1The fully automatic secondary thermal analyzer developed and produced by Beiten Tianpu Company fully meets the new standard.16 speciesTVOCSample testing requirements, with a very low background, good repeatability, all linear is better0.999and can work steadily.

     

     

     

     

     

        

     

     

     

     

    GB50325-2020TVOC16 component spectrum

    1, n-hexane 2. benzene 3. trichloroethylene 4. toluene 5. octene 6. butyl acetate 7. ethylbenzene 8. m., p-xylene 9. styrene 10. o-xylene 11. nonane 12. isooctanol. 13. n-undecane. 14. n-tetradecane 15. n-hexadecane

     

    Main applicable standards

    GB/50325-2020Standard for indoor environmental pollution control of civil building engineering

    GB/T18883-2020Indoor Air Quality Standards

    HJ/583-2010Ambient air-Determination of benzene series-Solid adsorption/thermal desorption-gas chromatography

    HJ/644-2013Ambient air-Determination of volatile organic compounds-Sorbent tube sampling-thermal desorption gas chromatography-mass spectrometry

    HJ/T400-2007Method for sampling and determination of volatile organic compounds and aldehydes and ketones in vehicles

    HJ/734-2014Stationary source emissions-Determination of volatile organic compounds by solid-phase adsorption-thermal desorption/gas chromatography-mass spectrometry

    GB/T33871-2017Total volatile organic compounds in toner (TVOC), determination of benzene and styrene

    GB 18587-2001Interior decoration materials Carpets, carpet pads and carpet adhesives Harmful substance release limit,GB 27630-2011Guide for Air Quality Assessment in Passenger Cars

    GB 36246-2018Synthetic material surface sports ground in primary and secondary schools

    IEC 62321-8-2017Determination of certain substances in electrotechnical products. Section8Part: Using gas chromatography mass spectrometry (GC-MS), with thermal cracking/Gas chromatography mass spectrometer with thermal desorption attachment (Py/TD-GC-MS) Determination of phthalate esters in polymers

    ASTM D7823-2016Using thermal desorption-gas chromatography/Standard Test Method for Determination of Low Level Modulated Phthalates in Polyvinyl Chloride Plastics by Mass Spectrometry

    ASTMD6196-2015Standard Practice for Selecting Sorbents, Sampling Parameters, and Thermal Desorption Analysis Conditions for Monitoring Volatile Organic Compounds in Air

    ASTMD7339-2012with a specific sorption tube and thermal desorption/Standard Test Method for Volatile Organic Compounds Evaporation from Carpets by Gas Chromatography

    BS PD CEN/TS 13649-2014Stationary source emissions-Specific gases-Determination of the mass concentration of organic compounds-Adsorption sampling method before solvent extraction or thermal desorption

  • Technical parameters:

    Adsorption tube temperature control range: room temperature-400℃, temperature control accuracy: ±1

    Six-way valve injection system temperature and control range: room temperature-220℃, temperature control accuracy: ±1

    Sample transfer tube temperature and control range: room temperature-220℃, temperature control accuracy: ±1

    focusing tube temperature control range:-40-450℃, heating rate>4800/min

    Cold trap temperature control range:-40-Room temperature, using electronic refrigeration device, temperature control accuracy: ±1

    Sample position:12Bit/24Bit

    Blowback flow:0100ml/min(continuously adjustable)

    Sample preparation flow:100ml/min

    Sample desorption, purge, injection and blowback time:0.0min999.9min

    Adsorption tube specifications: diameter:1/4inches, length:3.5Inches

    Power:800W

    Power:220V 50Hz

    Instrument size:610×350×510(mm)

    Instrument weight: approx.35kg

  • Main applicable standards

    GB/50325-2020Standard for indoor environmental pollution control of civil building engineering

    GB/T18883-2020Indoor Air Quality Standards

    HJ/583-2010Ambient air-Determination of benzene series-Solid adsorption/thermal desorption-gas chromatography

    HJ/644-2013Ambient air-Determination of volatile organic compounds-Sorbent tube sampling-thermal desorption gas chromatography-mass spectrometry

    HJ/T400-2007Method for sampling and determination of volatile organic compounds and aldehydes and ketones in vehicles

    HJ/734-2014Stationary source emissions-Determination of volatile organic compounds by solid-phase adsorption-thermal desorption/gas chromatography-mass spectrometry

    GB/T33871-2017Total volatile organic compounds in toner (TVOC), determination of benzene and styrene

    GB 18587-2001Interior decoration materials Carpets, carpet pads and carpet adhesives Harmful substance release limit,GB 27630-2011Guide for Air Quality Assessment in Passenger Cars

    GB 36246-2018Synthetic material surface sports ground in primary and secondary schools

    IEC 62321-8-2017Determination of certain substances in electrotechnical products. Section8Part: Using gas chromatography mass spectrometry (GC-MS), with thermal cracking/Gas chromatography mass spectrometer with thermal desorption attachment (Py/TD-GC-MS) Determination of phthalate esters in polymers

    ASTM D7823-2016Using thermal desorption-gas chromatography/Standard Test Method for Determination of Low Level Modulated Phthalates in Polyvinyl Chloride Plastics by Mass Spectrometry

    ASTMD6196-2015Standard Practice for Selecting Sorbents, Sampling Parameters, and Thermal Desorption Analysis Conditions for Monitoring Volatile Organic Compounds in Air

    ASTMD7339-2012with a specific sorption tube and thermal desorption/Standard Test Method for Volatile Organic Compounds Evaporation from Carpets by Gas Chromatography

    BS PD CEN/TS 13649-2014Stationary source emissions-Specific gases-Determination of the mass concentration of organic compounds-Adsorption sampling method before solvent extraction or thermal desorption

  • Main applicable standards

    GB/50325-2020Standard for indoor environmental pollution control of civil building engineering

    GB/T18883-2020Indoor Air Quality Standards

    HJ/583-2010Ambient air-Determination of benzene series-Solid adsorption/thermal desorption-gas chromatography

    HJ/644-2013Ambient air-Determination of volatile organic compounds-Sorbent tube sampling-thermal desorption gas chromatography-mass spectrometry

    HJ/T400-2007Method for sampling and determination of volatile organic compounds and aldehydes and ketones in vehicles

    HJ/734-2014Stationary source emissions-Determination of volatile organic compounds by solid-phase adsorption-thermal desorption/gas chromatography-mass spectrometry

    GB/T33871-2017Total volatile organic compounds in toner (TVOC), determination of benzene and styrene

    GB 18587-2001Interior decoration materials Carpets, carpet pads and carpet adhesives Harmful substance release limit,GB 27630-2011Guide for Air Quality Assessment in Passenger Cars

    GB 36246-2018Synthetic material surface sports ground in primary and secondary schools

    IEC 62321-8-2017Determination of certain substances in electrotechnical products. Section8Part: Using gas chromatography mass spectrometry (GC-MS), with thermal cracking/Gas chromatography mass spectrometer with thermal desorption attachment (Py/TD-GC-MS) Determination of phthalate esters in polymers

    ASTM D7823-2016Using thermal desorption-gas chromatography/Standard Test Method for Determination of Low Level Modulated Phthalates in Polyvinyl Chloride Plastics by Mass Spectrometry

    ASTMD6196-2015Standard Practice for Selecting Sorbents, Sampling Parameters, and Thermal Desorption Analysis Conditions for Monitoring Volatile Organic Compounds in Air

    ASTMD7339-2012with a specific sorption tube and thermal desorption/Standard Test Method for Volatile Organic Compounds Evaporation from Carpets by Gas Chromatography

    BS PD CEN/TS 13649-2014Stationary source emissions-Specific gases-Determination of the mass concentration of organic compounds-Adsorption sampling method before solvent extraction or thermal desorption